Portable Gas Detector Calibration Procedure

Portable (multi-gas) detectors used in confined spaces and maintenance work directly protect workers. Accuracy management is critical, yet a common misconception on site is that a bump test is enough. This guide from a KOLAS-accredited laboratory (no. KC23-416) walks through how portable detector calibration actually works and how it differs from a bump test.

Bump test ≠ calibration

PurposeConfirm the sensor responds and the alarm soundsAdjust readings to match standard gas concentration
Guarantees accuracyNoYes
WhenRoutine, before each usePeriodic (commonly 12 months)
CertificateNoneKOLAS-accredited certificate available

They complement rather than replace each other: bump test routinely to confirm function, calibrate periodically to restore accuracy.

How calibration proceeds

  • 1. Pre-check — housing, sensor condition, battery, and pump (for pumped units) are inspected. An expired sensor cannot be fixed by calibration and must be replaced first.
  • 2. Zero calibration — in clean air (or zero gas), the baseline is set so the reading shows 0 (20.9% for oxygen).
  • 3. Span calibration — certified standard gas of known concentration is applied and the sensitivity adjusted until the reading matches. This is the core of calibration.
  • 4. Verification and records — before/after readings and errors are recorded, and alarm setpoints verified.
  • 5. Certificate — a KOLAS-accredited certificate is issued. See the certificate guide for what it contains.

Courier calibration suits portables

Portables are small enough that courier calibration is usually the most convenient — inbound and on-site are also available. Check the calibration process, and send us the model, quantity, and target gas for a quick schedule and quote.

Frequently Asked Questions

No. A bump test only checks that the sensor responds to gas — it does not verify reading accuracy. Accuracy is guaranteed only by calibration against standard gas. Standard practice is routine bump tests plus periodic calibration.

As with fixed detectors, the user sets the interval — 12 months is the common recommendation. Portables face more shocks, movement, and temperature swings, so consider a shorter cycle under heavy use.

Within our KOLAS scope (gas analyzers), calibration is possible regardless of maker. Send us the model and target gas and we will confirm feasibility and schedule.

Yes. A KOLAS-accredited calibration certificate is issued on completion — see our certificate guide for what it contains.

Related Pages

  • ACE Electronics Co., Ltd.
  • Address: 192-26, Geurugogae-ro 221beon-gil, Gwangjeok-myeon, Yangju-si, Gyeonggi-do
  • Phone: +82-31-858-2020
  • Fax: +82-31-858-2022
  • Email: contact@acetop119.com

  • Terms of Service
  • Privacy Policy
  • Copyright © 1993-2026 ACE Electronics Co., Ltd. All Rights Reserved.